<- Back to Learning

What Are Mechanical Ventilator Waveforms?

Mechanical ventilator waveform (MVW): a line traced between sequential respiratory measurements collected by the ventilator during each breath.

Core idea: the ventilator continuously records data points over time. Those points are connected into graphical representations of respiratory behavior.

Clinical use: several waveforms are reviewed together because each waveform shows a different view of the same underlying dataset: the breath.

Primary waveform signals: modern ventilators commonly present flow, pressure, and volume as the main waveform views.

Scope: This page is about ventilator waveforms only. Waveforms are not ventilator modes. Modes are a separate discussion. Modes are classified from an evaluation of waveform behavior.

What is a Mechanical Ventilator

What is a Mechanical Ventilator?

A “mechanical ventilator,” or simply a “ventilator,” is an automated device that moves a controlled mixture of air and supplemental oxygen into and out of the lungs.

Key respiratory therapy terms: ventilation, gas exchange, oxygen delivery, lung movement, and respiratory failure.

The device supports patients who have stopped breathing or whose respiratory function has failed.

Definition adapted from Merriam-Webster.

Trends menu button on a mechanical ventilator screen

Trends view: historical ventilator information used to monitor disease severity, recovery, ventilation needs, and oxygenation requirements over time.

Real-Time view: live waveform information used to identify immediate changes such as coughing, breath stacking, and patient-ventilator asynchrony.

Lesson data: analog output ports were reconstructed into timestamped pressure, flow, volume, and breath-phase traces for playback and download.

Data Creation to Data Collection:
Mechanical ventilator data can be collected manually when a clinician or technician reviews the Trends and Real-Time displays at the bedside. The measurements behind those displays can also be collected automatically. In this lesson, waveform measurements came from the AVEA ventilator using the RS-232 interface and analog output signals.

Collection pathway: AVEA analog signals -> RS-232 connector -> Arduino Uno -> timestamped data stream -> workbook -> interactive waveform player on pages 12-19.

Trace

The diagram shows the path from AVEA analog output pins to Arduino capture, workbook conversion, and waveform plotting.

Export

The complete setup connected the AVEA ventilator output to the Arduino capture path so the analog output could become a timestamped workbook.

Prepare

A series of DuPont pins were inserted into the Arduino Analog Input Pins to receive RS-232 output signals.

Capture

The DuPont Pins were inserted into RS-232 pin 22, 23, 24, and 25. Data was collected polled as at 1 millisecond intervals.

Plot

The resulting values for every millisecond were exported as timestamped rows in a comma separated value field and plotted against clean AVEA waveform examples.

Reference: CareFusion. AVEA Ventilator System Service Manual, pp. 244-246 (Rev. D), Chapter 11: Appendices.

Reference: VIASYS Healthcare. AVEA Ventilator Systems Operator's Manual, Chapter 2: p. 2-23 and p. 2-24.

Workbook source: data.xlsx in the ZBZFirst/info repository.

Page reviewed and edited by Paul Michael Casillas, MSRC. Last reviewed: July 6, 2026. Last edited: July 16, 2026.